In which modes does ProPSV work?
The ventilation modes of ProPSV are CPAP, PSV, aPCV, PCV, ProSV and ProVT. The ventilation type is leak ventilation.
How does volume targeting work on ProPSV?
PSV, PCV and aPCV modes have a target volume feature. If the patient cannot reach the set target volume, ProPSV increases the pressure automatically.
The maximum pressure increase is set between 3 and 10 cmH2O; the feature can be switched off.
How long does ProPSV run on battery?
The internal lithium-ion battery of ProPSV lasts about 3 hours. ProPSV also runs from an external 12 V, 8 A DC source.
What data does ProPSV display and record?
The color LCD screen of ProPSV shows real-time pressure and flow graphs. Pressure, flow, tidal volume and respiratory rate can be monitored on the screen.
The SD card records up to 365 days of therapy data. The data is reported with SleepOne TrackPro; SleepOne Desktop is used for remote control and real-time graphs.
All alarms are saved in the memory.

What are the technical specifications of ProPSV?
| Specification | Value |
|---|---|
| Ventilation modes | CPAP, PSV, aPCV, PCV, ProSV, ProVT |
| Ventilation type | Leak ventilation |
| IPAP | 6-35 cmH2O |
| EPAP | 3-20 cmH2O |
| Target volume | 50-2000 ml |
| Respiratory rate | 4-50 bpm |
| I:E ratio | 1:0.3-1:4.0 |
| Inspiratory trigger | 1-5, Automatic |
| Expiratory trigger | 10-90% |
| Dimensions | 18.1 x 15.5 x 11.7 cm |
| Weight | 1.260 kg |
| Noise level | < 30 dBA |
| Humidification | SleepOne Heated Humidifier (optional) |
| Alarms | High pressure, low pressure, high volume, low volume, high respiratory rate, high leak, low battery, no SD card |
| Values monitored on the screen | Pressure, flow, tidal volume, respiratory rate |
| DC power supply | 12 V, 8 A |