Owgels OZ-5-01TW0 is listed in the catalogue as OZ-5-01 TWO.
What oxygen flow range does Owgels OZ-5-01TW0 operate in?
Owgels OZ-5-01TW0 delivers a continuous flow of oxygen between 0 and 5 L/min. Oxygen concentration is 93 ± 3%. The flow is set with the knob on top of the flow meter and read from the position of the black ball.
According to the user manual, Owgels OZ-5-01TW0 reaches the specified concentration within 2 minutes of operation at the set flow rate.
How does Owgels OZ-5-01TW0 produce oxygen?
Owgels OZ-5-01TW0 uses a pressure swing adsorption process and a molecular sieve to separate and concentrate oxygen from the air. Oxygen output pressure is 8 ± 2 psi and power consumption is 350 W.
In which environmental conditions does Owgels OZ-5-01TW0 operate?
Owgels OZ-5-01TW0 operates at 5-40 °C and 15-80% relative humidity. The operating altitude is 0-3000 m; above this altitude, oxygen concentration is insufficient.
According to the user manual, Owgels OZ-5-01TW0 is placed in a well-ventilated location at least 15-30 cm away from walls, furniture and curtains. The ventilation ports are at the bottom and back.

What is on the control panel of Owgels OZ-5-01TW0?
The display of Owgels OZ-5-01TW0 shows the total running time and alarm indicators. The panel has an on/off button and 4 indicator lights: low pressure, low concentration, power outage and power.

When does Owgels OZ-5-01TW0 give an alarm?
The alarms of Owgels OZ-5-01TW0 are medium priority; the indicator light flashes yellow every two seconds and an audible signal sounds:
- When the outlet pressure drops below 2.2 psi
- When the oxygen concentration drops below 82%
- When there is a power outage
- When start-up fails
- When there is no flow or low flow (on screen)
On Owgels OZ-5-01TW0, the alarm condition delay and the alarm signal generation delay are both 5 seconds.
How does the atomization outlet of Owgels OZ-5-01TW0 work?
Owgels OZ-5-01TW0 has a separate atomization outlet. Oxygen delivery and atomization cannot be performed at the same time.
The atomization pressure range of Owgels OZ-5-01TW0 is 70-170 kPa and the cup capacity is 8 ml. The user manual recommends cleaning the atomization components before each use.

When is Owgels OZ-5-01TW0 not to be used?
According to the user manual, Owgels OZ-5-01TW0 is not intended for life-supporting or life-sustaining situations or for an operating or surgical environment. It is also not intended for use with a non-adult population or with flammable anesthetics or flammable materials.
Owgels OZ-5-01TW0 is for use by prescription only. A backup oxygen source should be available in case of power outage or malfunction. Open flames should not be allowed within 3 m.
How often is Owgels OZ-5-01TW0 maintained?
The maintenance intervals in the Owgels OZ-5-01TW0 user manual are:
- Foam cabinet filter: cleaned weekly with warm water and mild detergent
- HEPA/compressor filter: lasts up to 3 years depending on the environment
- Final bacteria filter: inspected and replaced as necessary
- Atomization components: cleaned before each use
- Outer case: wiped with a cloth dampened with mild detergent
What are the technical specifications of Owgels OZ-5-01TW0?
| Specification | Value | Unit |
|---|---|---|
| Oxygen flow | 0-5 | L/min |
| Oxygen concentration | 93 ± 3 | % |
| Oxygen output pressure | 8 ± 2 | psi |
| Rated voltage / frequency | 220 V AC / 50 Hz | |
| Power consumption | 350 | W |
| Sound | ≤ 50 | dB(A) |
| Time to reach specified concentration | ≤ 2 | min |
| Weight | 19 | kg |
| Dimensions (W × D × H) | 34 × 30 × 65 | cm |
| Operating temperature | 5-40 | °C |
| Operating humidity | 15-80 | % RH |
| Operating altitude | 0-3000 | m |
| Storage and transport temperature | -20 to 60 | °C |
| Storage humidity | ≤ 80 | % RH |
| Electrical class | Class II, type BF | |
| Indicator lights | 4 | pcs |
| Atomization pressure range | 70-170 | kPa |
| Atomization cup capacity | 8 | ml |
| Shelf life | 60 | months |